SOME OBSERVATIONS ON DEHYDRATION AND DRYING OF ERYTHROCYTES FOR SCANNING ELECTRON MICROSCOPY

Main Article Content

F. Y.S. Haddad

Abstract

A comparative study on air drying of human erythrocytes for scanning electron microscopy (S E M), between acetone and ethanol for dehydration and drying was carried out, using a quick method. No morphological deformity or artifacts of erythrocytes was observed. More rapid air drying in acetone (3-5 seconds) was noticed as compared to ethanol (20-30 seconds). Acetone was recommended to be used for air drying, in the absence of critical point drying.

Downloads

Download data is not yet available.

Article Details

How to Cite
SOME OBSERVATIONS ON DEHYDRATION AND DRYING OF ERYTHROCYTES FOR SCANNING ELECTRON MICROSCOPY . (1992). The Iraqi Journal of Veterinary Medicine, 16(1), 42-47. https://doi.org/10.30539/ijvm.v16i1.1610
Section
Articles

How to Cite

SOME OBSERVATIONS ON DEHYDRATION AND DRYING OF ERYTHROCYTES FOR SCANNING ELECTRON MICROSCOPY . (1992). The Iraqi Journal of Veterinary Medicine, 16(1), 42-47. https://doi.org/10.30539/ijvm.v16i1.1610

References

Barber, V.C. and Boyde, A. (1968). Scanning electron microscopic studies on cilia. Z. Zellforsch. 84:269-284.

Fujita, T., Inoue, H. And Kodama, T. (1968). Scanning electron microscopy of the normal rheumatoid synovial membranes Arch. Hist. Jap. 29:511-522.

Fujita, T., Tokunaga, J. And Inoue, H. (1971). Atlas of scanning electron microscopy in medicine. Tokyo, Igaku Shoin Ltd.

Anderson, T.F. (1951). Techniques for the preservation of three dimensional structure in preparing specimens for the electron microscope. Trans. New York Acad. Sci. 13:130-134.

Boyde, A. And Wood, C. (1969). Preparation of animal tissues for surface scanning electron microscopy. J. Microscopy 90:221-249.

Al-Samarrai, S.F. (1975). Studies on preparation technique of human erythrocytes for clinical scanning electron microscopy. Arch. Histol. Jap., vol.38, No.4:321-334.

Al-Samarrai, S.F.(1986). Quick method for SEM. Abstract of the third Sceientific Congress of College of Medicine, University of Baghdad. April 15-17, pp. 42.

Al-Samarrai, S.F. (1987). Quick method for diagnostic S E M, Proceeding of the 24th Annual Medical Congress of the Iraqi Medical Society Baghdad. April 7-9, pp. 42.

Al-Samarrai, S.F.(1984). Critical point drying. Manual of Medical Electron Microscopy, pp. 26-27. Continuous Education College of Medicine, University of Baghdad.

Instruction Manual for Hitachi Model HCP-1 critical point drying apparatus. (1976), pp. 4.

Similar Articles